logo
Enviar mensaje
Inicio Noticias

Factors Affecting Accuracy of Barcol Durometer

Certificación
Porcelana Dongguan Quality Control Technology Co., Ltd. certificaciones
Porcelana Dongguan Quality Control Technology Co., Ltd. certificaciones
Comentarios de cliente
La pantalla táctil que el probador micro de la dureza de los vickers ha estado instalado hace unas semanas, todo es ACEPTABLE. Gracias por su consejo y servicio excelente.

—— Sr. David Chambers

Las gracias por la entrega rápida y nuestro cliente muy se satisfacen por el probador de la dureza de Rockwell, y esperan más órdenes pronto.

—— Neil Renato

Después de que visitáramos al cliente, finalmente decidían comprar el probador Brinell iBrin-413Z de la dureza, ahora nosotros han instalado la unidad en el sitio de cliente,

—— Fermin Lee

Gracias por los penetradores (diamante  cono 120 grados). Lo conseguimos recientemente y lo probó ya. Penetradores  sea bueno como esperamos. 

—— Dalius Skinulis

Las medidas son buenas para el probador Brinell HBST-3000.

—— Sun Chull Kim

Ey, el probador iVick-484ST de la dureza fue instalado la semana pasada y trabaja bien, las gracias por la ayuda profesional.

—— Sabinian Smith

Estoy alegre informarle que la instalación del probador Brinell de la dureza de SHB-3000C se ha acabado con éxito. ¡Quisiera agradecerle!

—— Robbin

Estoy en línea para chatear ahora
Compañía Noticias
Factors Affecting Accuracy of Barcol Durometer
últimas noticias de la compañía sobre Factors Affecting Accuracy of Barcol Durometer

 

Factors Affecting Accuracy of Barcol Durometer

 

Barcol hardness tester is widely used in the aluminum profile industry. In order to ensure the accuracy of the measurement results and the long-term stable operation of the instrument, users must strictly abide by the operating specifications and pay attention to relevant details during use.

 

The following is a further sorting and supplement of the precautions for using the Barcol hardness tester based on the information you provided:

 

1. Maintain the cleanliness of the sample

Before measurement, appropriate cleaning tools and methods (such as dust-free cloth, alcohol, etc.) should be used to remove dust, oil and other contaminants on the surface of the sample to ensure that the measuring point is clean and dry.

 

Avoid using cleaning tools that are too rough to avoid scratching the surface of the sample and affecting the measurement results.

 

2. Selection of measuring points

When selecting the measuring point, ensure that it is at least 5mm away from the edge of the sample to reduce the impact of the edge effect on the measurement results.

 

The distance between the two indentations should be greater than 6mm to avoid mutual interference and ensure that the measurement results of each indentation are independent and accurate.

 

For the same sample, multiple measuring points should be selected for testing as much as possible, and the average value should be taken as the final result to improve the reliability of the measurement.

 

3. Consider the influence of oxide film

For aluminum profiles with oxide layer, the influence of oxide film on hardness measurement results should be considered. If necessary, correction can be made by removing the oxide layer or by other methods.

 

For oxide film with known thickness, correction can be made by referring to the corresponding correction formula or standard value to obtain more accurate hardness value.

 

4. Standard operation

Calibration and correction: Before using the Barcol hardness tester, full scale calibration and calibration point calibration should be performed using the included Barcol hardness sheet to ensure the accuracy and reliability of the instrument.

 

Perform instrument calibration and maintenance regularly to ensure its long-term stable operation.

Measurement process: During the measurement, ensure that the surface of the specimen is perpendicular to the indenter and the bottom of the specimen is in full contact with the anvil to ensure the accuracy of the measurement results.

 

The grip force should be applied quickly and evenly to avoid low measurement values caused by slow force application.

Maintain the stability of the specimen and the instrument to avoid twisting or shaking during the measurement to prevent damage to the indenter and reduce measurement errors.

 

5. Other precautions

Avoid using the Barcol hardness tester under extreme temperature or humidity conditions to avoid affecting its measurement accuracy and stability.

 

For instruments that are not used for a long time, they should be stored in a dry, dust-free, non-corrosive gas environment and regularly maintained and serviced.

 

For operators, regular training and technical guidance should be provided to ensure that they can correctly use and maintain the Barcol hardness tester and improve the accuracy and reliability of the measurement results.

 